Voted Best Answer
Apr 06, 2021 - 08:58 AM
Yes, this could be used with a Fluke scopemeter. You will need to use the BB120 Banana to BNC adapters that were included with the 125b.
The Fluke will not have a way to convert the voltage output of the transducer to pressure so you will have to do that manually which is fairly easy since the WPS500 scales linearly by range. (i.e 1V = 1psi, 1V = 10psi, or 1V = 100psi depending on range setting)
The Fluke will not have a way to convert the voltage output of the transducer to pressure so you will have to do that manually which is fairly easy since the WPS500 scales linearly by range. (i.e 1V = 1psi, 1V = 10psi, or 1V = 100psi depending on range setting)
Source: SA