Jul 02, 2018 - 06:55 AM
The PV350 would be compatible with the Modis Edge as well.
However, due to a recent hardware change in the PV350, this transducer is no longer ideal for lab scope use.
In order to make the PV350 work better with multimeters, Fluke has changed the way the transducers outputs the signal. This produces a wavefrom as shown below. Any changes in pressure that occur between the 11-12hz sample points will not be displayed, creating a stepped appearance instead of a smooth graph.
However, due to a recent hardware change in the PV350, this transducer is no longer ideal for lab scope use.
In order to make the PV350 work better with multimeters, Fluke has changed the way the transducers outputs the signal. This produces a wavefrom as shown below. Any changes in pressure that occur between the 11-12hz sample points will not be displayed, creating a stepped appearance instead of a smooth graph.
Source: SA
Jul 02, 2018 - 06:57 AM
Due to a hardware change in the PV350, this transducer is no longer ideal for use with any lab scope.